Dew Point & Temperature Probe – Vaisala DMP5

Features

  • Measures humidity at temperatures up to +180 °C (+356 °F).
  • Dew point measurement range: −40 … +100 °C (−40 … +212 °F) Td/f.
  • Dew point measurement accuracy up to ±2 °C (±3.6 °F) Td/f.
  • Sensor purge provides superior chemical resistance.
  • Condensation-tolerant.
  • Modbus RTU over RS-485.
  • Compatible with Indigo transmitters and Insight PC software.
  • Traceable calibration certificate.
Enquiry
Category:

Dpstar is a leading supplier of environmental measurement sensors for temperature, relative humidity, pressure, dew point, CO2, and other environmental condition monitoring for industrial process application and research. Since the ’90s, we earned a reputation as a trusted partner to our clients through our commitment and dedication. We are an authorized distributor and supplier of measuring sensors solutions from renowned manufacturers Vaisala. Vaisala is a global leader in environmental and industrial measurement. Building on 80 years of experience, Vaisala provides observations for a better world. Vaisala is a reliable partner for customers around the world, offering a comprehensive range of innovative observation and measurement products and services.

Vaisala DRYCAP Dew Point and Temperature Probe DMP5 is designed for in-line humidity measurement in industrial drying applications with high temperatures and a wide dew point range. The long and robust steel probe and optional installation flange allow easy installation with adjustable depth to reach through insulation, for example in ovens.

DMP5 incorporates the Vaisala DRYCAP sensor, which is accurate, reliable, and stable. The sensor is condensation-resistant and immune to particulate contamination, oil vapor, and most chemicals. Sensor warming minimizes the risk of condensation accumulating on the sensor. If the DRYCAP sensor does get wet, it dries rapidly and recovers its swift response time. In low-humidity conditions, the sensor will auto-calibrate to ensure accurate measurement. DMP5 is part of the Indigo family and is compatible with the Indigo500 and Indigo200 series transmitters. It can be used with Vaisala Insight PC software for data visualization, easy probe configuration, and self-calibration.

Dimension

Key Benefits

Measure Dew Point Directly In Hot Processes
DMP5 is built for in-line dew point measurement in hot processes, where dew point can reach levels from -40°C (-40°F) all the way up to 100°C (212°F). The probe measures dew point with ± 2°C accuracy at temperatures of up to 140°C (284°F), but it can operate safely at temperatures up to 180°C (+365°F). Installation is simple, as the probe can be installed directly in the process. With in-line measurement, there is no need for a sampling system or sample line trace heating.

Vaisala DRYCAP Sensor For Accurate & Reliable Measurements
The Vaisala DRYCAP sensor is accurate, reliable, and stable. It is condensation-resistant and immune to particulate contamination, oil vapor, and most chemicals. Sensor warming minimizes the risk of condensation accumulating on the sensor. If the DRYCAP sensor does get wet, it dries rapidly and recovers. The unique auto-calibration functionality ensures accurate and stable measurement. The chemical purge function periodically heats the sensor to remove harmful chemicals. This helps to maintain measurement accuracy between calibration intervals in environments with high concentrations of chemicals and cleaning agents.

Use Standalone Or With An Indigo Transmitter
DMP5 is part of the Vaisala Indigo family, so it can also be used with Vaisala Indigo series transmitters. The transmitters offer various additional benefits such as a display for data visualization, easy access to probe configuration, and more extensive options for automation system connectivity, supply voltage, and wiring. The built-in Modbus RTU over RS-485 output means that the DMP5 can be used as a standalone device and can also be easily integrated into third-party systems. Whether using the DMP5 on its own or with an Indigo transmitter, it can be connected to the Vaisala Insight PC software for easy field calibration, probe configuration, and device analytics.

 

Specification

Measurement Performance
Dew point
Sensor DRYCAP 180S
Measurement range −40 … +100 °C (−40 … +212 °F) Td/f
Accuracy ±2 °C (±3.6 °F) Td/f
See accuracy graph
Response time 63 % [90 %] 1)
From dry to wet 5 s [10 s]
From wet to dry 45 s [5 min]
Temperature
Measurement range 0 … +180 °C (+32 … +356 °F) 2)
Accuracy at +100 °C (+212 °F) ±0.4 °C (±0.72 °F)
Temperature sensor Pt100 RTD Class F0.1 IEC 60751
Mixing ratio
Measurement range (typical) 0 … 1000 g/kg (0 … 7000 gr/lbs)
Accuracy (typical) ±12 % of reading
Absolute humidity
Measurement range 0 … 600 g/m3
Accuracy ±10 % of reading (typical)
1) Tested with sintered filter.
2) If sensor warming is enabled, temperature measurement is locked when humidity rises above 80 %RH and warming is switched on.
Operating Environment
Operating temperature range for probe head −40 … +180 °C (−40 … +356 °F)
Operating temperature range for probe body −40 … +80 °C (−40 … +176 °F)
Storage temperature −40 … +80 °C (−40 … +176 °F)
Measurement environment For air, nitrogen, hydrogen, argon, helium, and oxygen 1)
IP rating for probe body IP66
1) Consult Vaisala if other chemicals are present. Consider safety regulations with flammable gases.
Inputs & outputs
Operating voltage 15 … 30 V DC
Current consumption 10 mA typical, 500 mA max
Digital output RS-485, non-isolated
Protocols Modbus RTU
Output Parameters
Absolute humidity (g/m3) Relative humidity (%RH)
Absolute humidity at NTP (g/m3) Relative humidity (dew/frost) (%RH)
Dew point temperature (°C) Temperature (°C)
Dew/frost point temperature (°C) Water concentration (ppmv)
Dew/frost point temperature at 1 atm (°C) Water concentration (wet basis) (vol %)
Dew point temperature at 1 atm (°C) Water mass fraction (ppmw)
Dew point temperature (°C) Water vapor pressure (hPa)
Enthalpy (kJ/kg) Water vapor saturation pressure (hPa)
Mixing ratio (g/kg)
Compliance
EU directives EMC Directive (2014/30/EU)
RoHS Directive (2011/65/EU)
EMC compatibility EN 61326 1, industrial environment
Type approvals DNV GL certificate no. TAA00002YT
Compliance marks CE, China RoHS, RCM
Mechanical Specifications
Connector M12 5-pin A-coded male
Weight 436 g (15.37 oz)
Probe cable length 2 m (6.56 ft) or 10 m (32.8 ft)
Materials
Probe AISI 316L
Probe body AISI 316L
Cable jacket FEP

Download

Vaisala DMP5 Dew Point & Temperature Probe_DpstarGroup

Reviews

There are no reviews yet.

Be the first to review “Dew Point & Temperature Probe – Vaisala DMP5”

Your email address will not be published. Required fields are marked *